sqz sdf monitoring

Naoki, Sheila, Vicky remotely

We looked through the SQZ slow controls channels and saw that there were 30 unmonitored channels in the observe.snap, we've monitored many of them.  The two attached screenshots show the before and after list of unmonitored channels.  We've left unmonitored the sqz angle, and various channels that will be different lock to lock.  

This means that now, if the OPO pump ISS is turned off as it was last night (71497, 71493) we will be knocked out of observing.  Oli has also added a diag main message for this situation (71509), so the operators will get a notification of what has happened.  

In general, this problem may be caused by some issue that needs to be solved, like those that Vicky looked into this morning.  In general, the problem is that there is less light available for the pump than the ISS is trying to servo to, so we can get back to observing by reducing the power requested, following instructions in 70050.  This will require accepting some SDF diffs, and will probably result in a lower level of sqz than if the ISS was working correctly at it's normal power.  

For now, let's try this approach to this problem, that we get knocked out of observing, the operator gets a message from diag main, and tries to follow the instructions to reset the ISS and go back to observing.  The next day someone can look at what happened with the green power. 