Reports until 15:23, Saturday 22 July 2023
H1 DetChar (CAL, DetChar)
gabriele.vajente@LIGO.ORG - posted 15:23, Saturday 22 July 2023 (71614)
Noise between 18 and 21 Hz is due to calibration line upconversion by ASC motion

The DARM noise between 18 an 21 Hz is due to calibration line upconversion by ASC motion. 
This is shown by the bicoherence between DARM and ASC-DHARD_Y, and also by the fact thta DARM is coherent with calibration line signals multiplied by ASC signals (see two examples attached)

We should repeat a test with CAL lines off again now that the low frequency sensitivity is better.

Images attached to this report