I've recalibrated the PSL rotation stage this morning using the following steps:
- Took IMC offline and paused LASER_PWR guardian
- Ran the script /opt/rtcds/userapps/release/psl/h1/scripts/RotationStage/MoveRotationStage.py to create PSL_rotStageCalib_25Jul2023.txt (attached)
- Changed PSLvals in CalibRotStage.m for correct file name then ran to create calibration curve (attached) and get new fit values (in table)
- Input new fit values into calibration and accepted new values in SDF
- Searched for HOME on rotation stage
- Resumed LASER_PWR guardian to go back to 2W input power
| |
Power in (W) |
D |
B (Minimum power angle) |
C (Minimum power) |
| Old Values |
104.199 |
1.990 |
-24.816 |
0.000 |
| New Values |
100.917 |
1.990 |
-24.827 |
0.000 |