Displaying report 1-1 of 1.
Reports until 16:11, Monday 17 July 2023
H1 General (DetChar)
camilla.compton@LIGO.ORG - posted 16:11, Monday 17 July 2023 - last comment - 13:46, Tuesday 25 July 2023(71418)
H1:OAF-RANGE_BAND_9 edited to Moniter 52Hz noise

Marissa, Camilla

We edited the filters in H1:OAF-RANGE_RBP_9 to be a bandpass for the 52Hz noise (70950, 70250) we've been seeing. Hopefully it will be easier to monitor this peak. Can be monitored as H1:OAF-RANGE_BAND_9 from 22:50UTC. Filters attached.

Images attached to this report
Comments related to this report
ryan.crouch@LIGO.ORG - 17:14, Monday 17 July 2023 (71424)

Accepted in SDF

Images attached to this comment
camilla.compton@LIGO.ORG - 13:46, Tuesday 25 July 2023 (71692)

See attached plot of this channel over last few days. Some of this 52Hz noise seems to be coupled with the wind speed.

Images attached to this comment
Displaying report 1-1 of 1.