IFO is in OBSERVING as of 01:14 UTC
We lost lock at 23:54 UTC (alog) with no apparent cause
Lock Acquisition:
1. IR was not found on ALS_DIFF and so I moved the Diff Offset slider from 397 to 417 and IR caught immediately thereafter. No further touching was needed all the way to NLN. There was no wait between OMC_Whitening and NLN.
2. There were ASC SDF diffs (screenshot below), but they showed a difference of 0 (though were initially showing up as red). Hitting the "observe" intention too early on my part I think stopped guardian from clearing these because as soon as I toggled it back, all guardian nodes were ready for observing.
The sdf diff screen in the screenshot isn't the real diffs. This is showing "FULL TABLE" and "SORT ON SUBSTRING: DHARD_Y". How I left it yesterday. To see the actual diffs, you should change the dropdown boxes to "SETTING DIFFS" and "SHOW ALL". I expect the diffs were from waiting for the ADS camera's to converge.