More Oscillators to PCAL, New Oscillators to DARM Added; CAL_AWG_LINES Function Now Replaced by Newly Installed FE Oscillators
J. Kissel, D. Barker

As of today Dave helped me install the new front-end, EPICs controlled oscillators discussed in LHO:71746. Then, after crafting a few new MEDM screens (see comments below), I've turned ON some of those oscillators in order to replace the unstable function of the CAL_AWG_LINES guardian.

So, there're no "new" calibration lines (not since we turned CAL_AWG_LINES back ON last week at 2023-07-25 22:21:15 UTC -- see LHO:71706) -- but they're now driven by front-end, EPICs controlled oscillators rather than by guardian using the python bindings for awg (which was unstable across computer crashes, and other connection interruptions).

This is true as of the first observation segment today: 2023-08-01 22:02 UTC
However, due to a mishap with me misunderstanding the state of the PCALY SDF system (see LHO:71879), I accidentally overwrote the PCALXY comparison line at 284.01 Hz, and we went into observe. Thus, 
The short observation segment between 22:02 - 22:11 UTC is out of nominal configration, because there's no PCALY line contributing to the PCALXY comparison.
 
The was rectified by the second observation segment starting on 2023-08-01 22:16 UTC.

Also, because of these changes the subtraction team should switch their witness channel for the DARM_EXC frequencies to H1:LSC-CAL_LINE_SUM_DQ.
The PCALY witness channel remains the same, H1:CAL-PCALY_EXC_SUM_DQ, as the newly used oscillators sum in to the same channel.
Below, I define which oscillator number is assigned to which frequency.

Here's the latest list of calibration lines:
Freq (Hz)   Actuator                   Purpose                      Channel that defines Freq             Changes Since Last Update (LHO:69736)     
8.825       DARM (via ETMX L1,L2,L3)   Live DARM OLGTFs             H1:LSC-DARMOSC1_OSC_FREQ              Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
8.925       PCALY                      Live Sensing Function        H1:CAL-PCALY_PCALOSC5_OSC_FREQ        Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
11.475      DARM (via ETMX L1,L2,L3)   Live DARM OLGTFs             H1:LSC-DARMOSC2_OSC_FREQ              Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
11.575      PCALY                      Live Sensing Function        H1:CAL-PCALY_PCALOSC6_OSC_FREQ        Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
15.175      DARM (via ETMX L1,L2,L3)   Live DARM OLGTFs             H1:LSC-DARMOSC3_OSC_FREQ              Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
15.275      PCALY                      Live Sensing Function        H1:CAL-PCALY_PCALOSC7_OSC_FREQ        Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
24.400      DARM (via ETMX L1,L2,L3)   Live DARM OLGTFs             H1:LSC-DARMOSC4_OSC_FREQ              Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
24.500      PCALY                      Live Sensing Function        H1:CAL-PCALY_PCALOSC8_OSC_FREQ        Former CAL_AWG_LINE, now driven by FE OSC; THIS aLOG
15.6        ETMX UIM (L1) SUS          \kappa_UIM excitation        H1:SUS-ETMY_L1_CAL_LINE_FREQ          No change
16.4        ETMX PUM (L2) SUS          \kappa_PUM excitation        H1:SUS-ETMY_L2_CAL_LINE_FREQ          No change
17.1        PCALY                      actuator kappa reference     H1:CAL-PCALY_PCALOSC1_OSC_FREQ        No change
17.6        ETMX TST (L3) SUS          \kappa_TST excitation        H1:SUS-ETMY_L3_CAL_LINE_FREQ          No change
33.43       PCALX                      Systematic error lines       H1:CAL-PCALX_PCALOSC4_OSC_FREQ        No change
53.67         |                            |                        H1:CAL-PCALX_PCALOSC5_OSC_FREQ        No change
77.73         |                            |                        H1:CAL-PCALX_PCALOSC6_OSC_FREQ        No change
102.13        |                            |                        H1:CAL-PCALX_PCALOSC7_OSC_FREQ        No change
283.91        V                            V                        H1:CAL-PCALX_PCALOSC8_OSC_FREQ        No change
284.01      PCALY                      PCALXY comparison            H1:CAL-PCALY_PCALOSC4_OSC_FREQ        Off briefly between 2023-08-01 22:02 - 22:11 UTC, back on as of 22:16 UTC
410.3       PCALY                      f_cc and kappa_C             H1:CAL-PCALY_PCALOSC2_OSC_FREQ        No Change
1083.7      PCALY                      f_cc and kappa_C monitor     H1:CAL-PCALY_PCALOSC3_OSC_FREQ        No Change
n*500+1.3   PCALX                      Systematic error lines       H1:CAL-PCALX_PCALOSC1_OSC_FREQ        No Change (n=[2,3,4,5,6,7,8])
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:31, Tuesday 01 August 2023 (71884)GRD
As a part of depricating CAL_AWG_LINES, I've updated the ISC_LOCK guardian to use the new main switches for the DARM_EXC lines for the transitions between NOMINAL_LOW_NOISE and NLN_CAL_MEAS. 

That main switch channel is H1:LSC-DARMOSC_SUM_ON, which enables excitations to flow through to the DARM error point when set to 1.0 (and blocks it when set to 0.0).

I've committed the new version of ISC_LOCK to the userapps repo, rev 26039.
jeffrey.kissel@LIGO.ORG - 15:36, Tuesday 01 August 2023 (71885)
Here's the updated 
/opt/rtcds/userapps/release/lsc/common/medm/
    LSC_OVERVIEW.adl
    LSC_DARM_EXC_OSC_OVERVIEW.adl
    LSC_CUST_DARMOSC_SUM_MTRX.adl

The new DARM oscillators screen (LSC_DARM_EXC_OSC_OVERVIEW.adl) is linked in the top-middle of the LSC_OVERVIEW.adl. The only sub screen on the LSC_DARM_EXC_OSC_OVERVIEW.adl is the summation matrix (LSC_CUST_DARMOSC_SUM_MTRX.adl).

I have not yet gotten to adding all the new PCAL oscillators to their MEDM screens, but I'll do so in the fullness of time.
Images attached to this comment
ansel.neunzert@LIGO.ORG - 15:40, Monday 21 August 2023 (72358)

detchar-request git issue for tracking purposes.

jeffrey.kissel@LIGO.ORG - 08:35, Tuesday 29 August 2023 (72503)
I found a bug in the
    /opt/rtcds/userapps/release/lsc/common/medm/
        LSC_DARM_EXC_OSC_OVERVIEW.adl
where DARMOSC1's TRAMP field was errantly displayed as all the 10 oscillator's TRAMPs; a residual from the copy pasta I made during the screen generation.

Fixed it. Now committed to the above location as of rev 26170.
jeffrey.kissel@LIGO.ORG - 16:51, Tuesday 29 August 2023 (72532)
Finally got around to updating the PCAL screens. Check out 
    /opt/rtcds/userapps/release/cal/common/medm/
        PCAL_END_EXC.adl
        CAL_PCAL_OSC_SUM_MATRIX.adl
as of userapps repo rev 26179.

See attached screenshots.
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 14:49, Thursday 07 September 2023 (72742)
This design is the result of ECR E2300227 and IIET:28681.