Reports until 20:13, Tuesday 23 July 2013
H1 ISC
jeffrey.kissel@LIGO.ORG - posted 20:13, Tuesday 23 July 2013 - last comment - 18:07, Wednesday 24 July 2013(7194)
State of IFO For Tonight
For the DetCharians: as of
1058670016
July 23 2013 20:00 PDT
Jul 24 2013 03:00:00 UTC
we are leaving the IFO in the following configuration:

Input Mode Cleaner: LOCKED (offload to MC2 stages M2 and M2 only, under slow WFS control)
Green Arm: LOCKED (no offload to HEPI)

HAM2 & HAM3 HEPI: Floating, with offsets, but no control
HAM2 & HAM3 ISI: Level 2 isolated
All HSTS SUS: Level 1.5 "resg" damping filters filter
PR3: Level 1.0 damping filters

ITM HEPI: Locked
ITM ISI: Level 2 isolated
ITM QUAD: Level 2.1 damped

BS HEPI: Level 2 isolated 
BS ISI: Level 2 isolated, but had to diable GS13 watchdogs while isolating to get it up and running
BS Triple: Level 2.0 damping filters

ETM HEPI: Level 2 isolated
ETM ISI: Level 2 isolated
ETM QUAD: Level 2.1 damped

May it stay locked all night!
The calibrated channel for green is H1:ALS-Y_REFL_CTRL_OUT_DQ, in [Hz] (so multiply by lambda*(L/c) to get [m])
The calibrated channels for the IMC are 
IMC_L = H1:IMC-X_DQ in [m]
IMC_F = H1:IMC-F_OUT_DQ in [m]


Comments related to this report
jeffrey.kissel@LIGO.ORG - 18:07, Wednesday 24 July 2013 (7212)
Note, I missed Vincent's comment about turning on the ST0 to ST1 feed forward in his BSC-ISI and BSC-HPI instruction manual, so none of the BSC-ISIs had their ST0-ST1 FF on for this lock stretch. (This may explain the difficulty behind getting ISI-BS up and running).