Closes FAMIS #25490
ISS diff power is currently sitting around 3.3% and has been sitting around that for the past 18 hours (see attachment) (minus a dip between 15-14 hours ago - presumeably related to the saturation mentioned below).
Laser Status:
NPRO output power is 1.826W (nominal ~2W)
AMP1 output power is 67.19W (nominal ~70W)
AMP2 output power is 134.8W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PMC:
It has been locked 3 days, 0 hr 16 minutes
Reflected power = 16.23W
Transmitted power = 107.9W
PowerSum = 124.2W
FSS:
It has been locked for 0 days 19 hr and 15 min
TPD[V] = 0.9439V
ISS:
The diffracted power is around 3.3%
Last saturation event was 0 days 15 hours and 11 minutes ago
Possible Issues:
ISS diffracted power is high