H1 was out of observing from 08:04 UTC to 08:41 UTC because the SQZ ISS had saturated. Following the instructions in alog 70050, I was able to recover the squeezer and resume observing. The opo_grTrans_setpoint_uW within sqzparams.py is currently set to 50, and attached is the SDF diff I accepted for the adjusted OPO TEC temperature.
H1 back observing at 151Mpc.