pyDARM and the GDS pipeline (which produces H1:GDS-CALIB_STRAIN
) relies on exports of the CAL-CS Foton filterbanks installed on the front end. The filters installed on the front end and those used by the calibration pipeline need to be in sync to properly calibrate the IFO.
I'm attaching a comparison of the inverse sensing and ETMX actuation Foton tfs as installed in the CAL-CS path (plots). The inverse sensing, ETMX UIM, and ETMX PUM tf exports match well between what is currently installed on the front end and what pyDARM thinks is installed. That's good. However, the ETMX TST stage has a significant deviation above 10Hz. This can be seen on the last page of the attached PDF, which shows both tfs overlaid on the same bode plot (left column) and their residual, also in the form of a bode plot (right column).
This discrepancy is not ideal and should be fixed by re-exporting the TST foton export for pyDARM to use it and be included in the next calibration export at LHO as per LHO:69563.
T2000022 is a good resource with instructions on how to export the appropriate TFs from the CAL-CS foton banks.