J. Kissel As a result of recent findings that the collection of lines used to probe the low frequency sensing function at PCALY_EXC 8.925, 11.575, 15.275, 24.5 [Hz] DARM_EXC 8.825, 11.475, 15.175, 24.4 [Hz] are causing excess non-linearities found when looking over long time scales --- see LHO:71964 --- I've turned OFF these calibration lines as of 2023-08-09 16:40 UTC. We were already out of observing for PEM injections anyways, and Robert relinquished a moment for me to quickly turn these off. I've accepted the changes in both the PCALY (h1caley) and OMC (h1omc) front-end model's SDFs against both the safe and OBSERVE.snaps such that this sticks. Here's the latest list of calibration lines: Freq (Hz) Actuator Purpose Channel that defines Freq Changes Since Last Update (LHO:69736) 8.825 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC1_OSC_FREQ Turned OFF; THIS aLOG 8.925 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC5_OSC_FREQ Turned OFF; THIS aLOG 11.475 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC2_OSC_FREQ Turned OFF; THIS aLOG 11.575 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC6_OSC_FREQ Turned OFF; THIS aLOG 15.175 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC3_OSC_FREQ Turned OFF; THIS aLOG 15.275 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC7_OSC_FREQ Turned OFF; THIS aLOG 24.400 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC4_OSC_FREQ Turned OFF; THIS aLOG 24.500 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC8_OSC_FREQ Turned OFF; THIS aLOG 15.6 ETMX UIM (L1) SUS \kappa_UIM excitation H1:SUS-ETMY_L1_CAL_LINE_FREQ No change 16.4 ETMX PUM (L2) SUS \kappa_PUM excitation H1:SUS-ETMY_L2_CAL_LINE_FREQ No change 17.1 PCALY actuator kappa reference H1:CAL-PCALY_PCALOSC1_OSC_FREQ No change 17.6 ETMX TST (L3) SUS \kappa_TST excitation H1:SUS-ETMY_L3_CAL_LINE_FREQ No change 33.43 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC4_OSC_FREQ No change 53.67 | | H1:CAL-PCALX_PCALOSC5_OSC_FREQ No change 77.73 | | H1:CAL-PCALX_PCALOSC6_OSC_FREQ No change 102.13 | | H1:CAL-PCALX_PCALOSC7_OSC_FREQ No change 283.91 V V H1:CAL-PCALX_PCALOSC8_OSC_FREQ No change 284.01 PCALY PCALXY comparison H1:CAL-PCALY_PCALOSC4_OSC_FREQ Off briefly between 2023-08-01 22:02 - 22:11 UTC, back on as of 22:16 UTC 410.3 PCALY f_cc and kappa_C H1:CAL-PCALY_PCALOSC2_OSC_FREQ No Change 1083.7 PCALY f_cc and kappa_C monitor H1:CAL-PCALY_PCALOSC3_OSC_FREQ No Change n*500+1.3 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC1_OSC_FREQ No Change (n=[2,3,4,5,6,7,8])