H1 AOS (DetChar)
robert.schofield@LIGO.ORG - posted 18:47, Wednesday 09 August 2023 (72118)
ETMX bias that minimizes grounding noise coupling hasnt changed much since January, and PEM injections at 60W

The changes in environmental coupling with the change from 75W to 60W ( https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=70808 ) require that we re-measure some of the coupling functions that were measured in the pre-run set of PEM injections in May. Today I began those injections by injecting acoustically in the LVEA, as well as shaking HAM6 and the ITMY cryopump. I also set up for acoustic injections at EX.

In addition Sheila and I did some bias manipulation for ETMX. I injected currents onto the EX building ground and swept the ETMX bias as I had in January (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=67075). The figure shows that there has not been much change in the bias setting that minimizes grounding noise.

 