Reports until 01:03, Thursday 25 July 2013
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 01:03, Thursday 25 July 2013 - last comment - 02:37, Thursday 25 July 2013(7216)
Infrared locked for the first time to the arm cavity

[Stefan and Kiwamu]

 We handed the CARM control from the green ALS beatnote sensor over to the REFL infrared demodulated signal. This worked well.

In this configuration, the PSL frequency was locked to the arm cavity and no green light was involved any more. The lock stayed for more than 20 minutes and was long enough to do some further noise investigations.

 

Update on Noise plot:

We tested two configurations :

In both cases we have three sensors to evaluate the noise performance. We had two channels for the ALS beatnote, which are REFL_SERVO_SLOW and ALS-C_COMM_A_RF_I. The difference is that COMM_A_RF_I is digitized before the common mode board and REFL_SERVO is digitized after the board. Ideally they should be always the same. The last signal is the infrared signal, REFLAIR_A_RF9_I, which is the one derived from the reflected light off of the arm cavity and demodulated at 9 MHz.

 

Configuration (A)

(Red): The infrared sensor. Therefore this is the in-loop signal for CARM

(Blue): Beat signal at ALS-C_COMM_A_RF_I. Out-of-loop.

(Green) : Beat signal at LSC-REFL_SERVO. Out-of-loop.

Configuration (B)

(Brown): The infrared sensor. Out-of-loop

(Pink): Beatnote at ALS-C_COMM_A_RF_I

(Cyan): Beatnote at LSC-REFL_SERVO. This is the in-loop error signal

Some other noise

(Black): beat-note readout electronics noise at LSC-REFL_SERVO

 

What we did:

Images attached to this report
Non-image files attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 01:14, Thursday 25 July 2013 (7218)
Attached is the attempt to calibrate MC2 using the beat-node signal. The horizontal bar is at 6e-16m/ct.
The variability is most likely due to saturation. Without frequency feed-back the beat-node signal tends to run into its range limit.

The calibration should be redone with the REFLAIR diode.
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 01:17, Thursday 25 July 2013 (7219)
Here's the plot I'd made to re-create this transfer function in the model, as requested by Kiwamu and Stefan.
Non-image files attached to this comment
stefan.ballmer@LIGO.ORG - 02:37, Thursday 25 July 2013 (7222)
Some additional items:
- the REFLAIR_A_RF9_I was calibrated into Red Hz by adding a filter with a gain of 82Hz/(2*1000cts) and a z82:p1e3 cavity pole compensator.
- We noticed the the green PDH error sigal at EY (as seen at the analog Imon signal from the demos board) has about 2.5Vpkk signal at ~28kHz. Given that it sometimes bleeds over into the Q sigal, I am pretty sure that we are saturating an RF amplifier before the demos board.