PSL Report FAMIS report:
Laser Status:
NPRO output power is 1.829W (nominal ~2W)
AMP1 output power is 67.15W (nominal ~70W)
AMP2 output power is 135.5W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PMC:
It has been locked 20 days, 6 hr 1 minutes
Reflected power = 16.33W
Transmitted power = 109.2W
PowerSum = 125.5W
FSS:
It has been locked for 0 days 10 hr and 48 min
TPD[V] = 0.9056V
ISS:
The diffracted power is around 2.4%
Last saturation event was 0 days 10 hours and 48 minutes ago
Possible Issues: None