louis.dartez@LIGO.ORG - posted 17:15, Thursday 31 August 2023 (72531)
LHO cal epochs since start of O4
As we are gearing up to regenerate our hourly uncertainty estimates spanning back to the start of O4, I needed to go back through the history of LHO calibration changes, the IFO changes they address, and the identification and placement of each new epoch for the sensing and actuation functions. The table below lists the calibration epochs since the start of O4. For each epoch, I include some relevant alogs that chronicle which IFO changes triggered each epoch. More in depth information on each change to the LHO calibration pipeline can be found in the living Record of Real-Time Calibration Pipeline Parameter Changes (DCC: T2300297).

Date Epoch Type Reason for new epoch
20230504T055052Z Sensing, Actuation new SRCL offset (LHO:69289), TCS changes (LHO:69032), and PCALX cal line change (LHO:69303). More in LHO:69332. This report was *not* exported to the front end.
20230506T182203Z Sensing Changed SRCL offset, DARM loop. More in LHO:69561. N.B. This report marks a new epoch for the sensing function but it did not get exported to the front end (i.e. the "calibration" was not updated using this report). Instead, the report that was used to update the calibration and includes the IFO changes listed in LHO:69561 is 20230510T062635Z.
20230621T191615Z Sensing Moved back to 60W input power(LHO:70693). 20230621T191615Z marks the start of the epoch but the report used to update the IFO calibration is 20230621T211522Z.
20230628T015112Z Sensing OM2 was heated (first time). See LHO:70849.
20230716T034950Z Sensing OM2 was unheated. See LHO:72524.
20230727T162112Z Sensing OM2 was reheated. See LHO:72523.
20230817T214248Z Actuation 3.2kHz filter restored in L3 actuation path. See LHO:72043
Several of the calibration reports had not been properly tagged yet to reflect the epoch boundaries defined above. I manually went through each one and made sure to do so, but this necessitated the reprocessing/regeneration of several calibration reports to ensure that the GPR calculations for the sensing and actuation functions only included the relevant set of measurements. Here is a list of the reports that were reprocessed with the appropriate epoch boundaries. Before being overwritten each report was backed up to /ligo/groups/cal/H1/reports/archive/. 06/21 reason: 60W 06/27 14:53 : om2 hot 20230628T015112Z [regenerated for epoch-sensing] 07/12 16:05 : om2 cold 20230716T034950Z [regenerated for epoch-sensing] 07/19 08:15 : om2 hot 20230727T162112Z [regenerated for epoch-sensing] 20230802T000812Z [regenerated for new epoch-sensing] 20230817T214248Z [regenerated for epoch-actuation] 20230823T213958Z [regenerated for new epoch-sensing and new epoch-actuation] actuation: 05/04: start of O4 08/08: 3.2kHz filter 20230817T214248Z [regenerated for epoch-actuation]