J. Kissel, L. Dartez After the calibration update today (2023-08-31; LHO:72594) we now have automatically generated comparisons between measured systematic error in the GDS-CALIB_STRAIN channel (as measured directly by PCAL) vs the traditional model of the systematic error. The results look great. I attach two automated comparisons to highlight the difference: Before from 1377445828 2023-08-30 15:50 UTC: - pre- 2023-08-30 maintenance, - pre- DARM2 FM8 boost turn on, - \kappa_T is still a 1.08, so DELTAL_EXTERNAL (which isn't corrected for time-dependence) disagrees with GDS-CALIB_STRAIN which does. - still not having updated the sensing function since we turned OM2 TSAMS back on - the modeled systematic error is still using GPR fits of measured data from 2023-06-21 and prior vs. After from 1377561032 2023-08-31 23:50 UTC: - post-calibration update - DARM2 FM8 boost is accounted for, - DELTAL_EXTERNAL and GDS-CALIB_STRAIN agree - All TDCFs are close to 1.0 - Modeled systematic error is now using GPR fits are now using an updated collection of measurements.