The ITMY ISI watchdog tripped, hasn't shown up on H1locklosses yet.
We're going through an initial alignment on the relock.
Reaquired NLN at 13:17, just waiting on the camera_servo
ITMY IST ST2 watchdog tripped at 13:21, not totally sure why. We lost lock 4 minutes later
These lock losses were caused by the ITMY capacitive position sensors glitching. See LHO:72683.