Reports until 08:26, Saturday 16 September 2023
H1 PSL
oli.patane@LIGO.ORG - posted 08:26, Saturday 16 September 2023 (72910)
PSL Weekly FAMIS

Closes FAMIS#26209, last completed Sept 8th


Laser Status:
    NPRO output power is 1.833W (nominal ~2W)
    AMP1 output power is 67.19W (nominal ~70W)
    AMP2 output power is 134.8W (nominal 135-140W)
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN

PMC:
    It has been locked 41 days, 0 hr 12 minutes
    Reflected power = 16.6W
    Transmitted power = 109.0W
    PowerSum = 125.6W

FSS:
    It has been locked for 0 days 19 hr and 29 min
    TPD[V] = 0.8165V

ISS:
    The diffracted power is around 2.4%
    Last saturation event was 0 days 19 hours and 29 minutes ago


Possible Issues: None