Power Supply Health Check

Per WP11435 I performed a health check on the Kepco power supplies at EY, EX, and CER.

EY -All supplies pass, no excess vibrations, no sounds, good airflow, no temperature issues.

EX - All supplies pass, no excess vibrations, no sounds, good airflow, no temperature issues.

CER - All but one supplies pass, no excess vibrations, no sounds, good airflow, no temperature issues.

** ISC C1 - Slot 38 RHS N18V 6A 99F at fan housing, 105 at rear, very slight fan stumbling / rumbling.  We need to keep an eye on this one, spare supply is staged.

While heading out to EX I spied some wildlife very close to the OSB.  Coyote attacks on people are very rare.  What to do if a coyote approaches you, make noise, clap your hands, whistle, they will move along.