ETMX OL values, gains and offsets

I measured the ETMX OL values

prettyOSEMgains('H1','ETMX')

M0F1 19894 1.508  -9947
M0F2 29453 1.019 -14727
M0F3 30661 0.978 -15330
M0LF 24512 1.224 -12256
M0RT 24724 1.213 -12362
M0SD 21784 1.377 -10892
R0F1 28922 1.037 -14461
R0F2 23013 1.304 -11507
R0F3 25716 1.167 -12858
R0LF 26662 1.125 -13331
R0RT 24388 1.230 -12194
R0SD 21961 1.366 -10980
L1UL 24267 1.236 -12133
L1LL 26538 1.130 -13269
L1UR 24545 1.222 -12273
L1LR 26259 1.142 -13130
L2UL 17935 1.673  -8967
L2LL 18726 1.602  -9363
L2UR 25124 1.194 -12562
L2LR 25518 1.176 -12759

Comments related to this report
mark.barton@LIGO.ORG - 15:15, Tuesday 30 July 2013 (7288)

I entered the gains and offsets into the OSEMINF screens and updated and committed the /opt/rtcds/userapps/release/sus/h1/burtfiles/h1susetmx_safe.snap .

mark.barton@LIGO.ORG - 12:37, Wednesday 31 July 2013 (7301)

Betsy was concerned about the M0 F1 and SD OL values, which were a bit low (in spec, but rather lower than average). So she swapped out those two and I remeasured those two values only. The new master list is as follows:

M0F1 24018 1.249 -12009 (new)
M0F2 29453 1.019 -14727
M0F3 30661 0.978 -15330
M0LF 24512 1.224 -12256
M0RT 24724 1.213 -12362
M0SD 28519 1.052 -14259 (new)
R0F1 28922 1.037 -14461
R0F2 23013 1.304 -11507
R0F3 25716 1.167 -12858
R0LF 26662 1.125 -13331
R0RT 24388 1.230 -12194
R0SD 21961 1.366 -10980
L1UL 24267 1.236 -12133
L1LL 26538 1.130 -13269
L1UR 24545 1.222 -12273
L1LR 26259 1.142 -13130
L2UL 17935 1.673  -8967
L2LL 18726 1.602  -9363
L2UR 25124 1.194 -12562
L2LR 25518 1.176 -12759

I entered the new values into the OSEMINF screen and redid the safe.snap.

betsy.weaver@LIGO.ORG - 16:02, Wednesday 31 July 2013 (7307)

Now with associated serial numbers which are also in ICS:

M0F1 - 643

M0F2 - 486

M0F3 - 414

M0LF - 580

M0RT - 507

M0SD - 563

R0F1 - 426

R0F2 - 497

R0F3 - 504

R0LF - 468

R0RT - 418

R0SD - 434