Since the EX system was added to the DAQ we have been seeing two problems in the frame:
Alex found that there is a limit on the number of DCUs acquired by the DAQ, the limit is 85. We had 94 DCUs in the DAQ.
We removed 16 DCUs from the DAQ this afternoon, and both problems have gone away (Patrick verified all DUST channels have good data).
Alex will test a fix to allow the full complement of DCUs in the DAQ next week. For now the systems listed below are out of the DAQ:
h1iopsusauxh56, h1susauxh56
h1iopsusquadtst, h1susquadtst
h1iopsusbstst, h1susbstst
h1iopseitst, h1isitst
h1iopsusex, h1iopseiex, h1iopiscex, h1iopsusauxex
h1iopsusey, h1iopseiey, h1iopiscey, h1iopsusuxey