Displaying report 1-1 of 1.
Reports until 00:22, Saturday 30 September 2023
H1 General (SUS)
anthony.sanchez@LIGO.ORG - posted 00:22, Saturday 30 September 2023 - last comment - 09:19, Saturday 30 September 2023(73178)
Friday Ops Eve Shift End

TITLE: 09/30 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 153Mpc
INCOMING OPERATOR: Oli
SHIFT SUMMARY:

Uknown Lockloss: 00:05 UTC
relocking started at 00:06 UTC

Nominal_LOW_NOISE reached and PI Mode 31 Spiked to 150! at 00:58 UTC

ITMY Mode 5 is increasing steadily since the beggining of the lock 3:08 UTC I changed the Damping setting from -0.04 to 0 to +0.04 to see If I could damp it. Channel name : H1:SUS-ITMY_L2_DAMP_MODE5_GAIN
3:31 UTC +0.04 gain is taking ITMY mode 5 back down.
The issue is that ITMY Mode 5 is coupled with ITMY mode 6 and while I have changed the gain setting for Mode5 and now mode 5 is slowly going down. Unfortunately Mode 6 is still ringing up and has been for the last 3 locks.
Next shift should keep an eye on those Violin modes. Plots for the mode 5 & mode 6  over the last 2 days attached.
Tagging SUS.
 

LOG:
no log

 

Images attached to this report
Comments related to this report
rahul.kumar@LIGO.ORG - 09:19, Saturday 30 September 2023 (73185)SUS

ITMY mode 05/06 settings are tricky, hence if they are rising it is safer to switch OFF the gain (or trying reducing it by 50%) and then see how it goes. Sign change can make it angry as we are seeing it right now (see alog 73180)