Reports until 19:03, Wednesday 18 October 2023
H1 SQZ
victoriaa.xu@LIGO.ORG - posted 19:03, Wednesday 18 October 2023 - last comment - 15:29, Monday 20 November 2023(73562)
NLG HD sweep, HD mystery losses fully resolved, ~8dB HD SQZ with new crystal spot

Naoki, Sheila, Camilla, Vicky

Summary: After yesterday's crystal move LHO:73535, we re-aligned SQZT7, and now see 8 dB SQZ on the homodyne, up to measured NLG=114 without a phase noise turnaround! This fully resolves the homodyne loss budget, there is 0 mystery loss remaining on the homodyne, from which we can infer 0 mystery losses in HAM7. Back to the IFO afterwards, after 1 day at this new crystal spot, squeezing in DARM is about 4.5dB - 4.8dB, reaching almost 5dB at the start of lock.

We first re-aligned the homodyne to the IFO SQZ alignment, which reached 4.8dB SQZ in DARM yesterday, so we are more confident the alignment back through the VOPO is not clipping. In yesterday's measurements, we had a sign error in the FC-ASC offloading script, which brought us to a bad alignment with limited homodyne squeezing, despite high 98% fringe visibilities. Attached is a screenshot of homodyne FC/ZM slider values with FC+SQZ ASC's fully offloaded (correctly), to which the on-table SQZT7 homodyne is now well-aligned. After Sheila re-aligned the homodyne to the screenshotted FC/ZM values, fringe visibilities are PD1 = 98.5% (loss 3.1%), PD2 = 97.8% (loss 4.2%).

We then did an NLG sweep on the homodyne, from NLG=2.4 (opo trans 20uW) to NLG=114 (opo trans 120uW). Measurements below and attached as .txt, DTT is attached, plots to follow.

-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

unamplified_ir = 0.0014 (H1:SQZ-OPO_IR_PD_LF_OUT_DQ with pump shuttered)

NLG = amplified / unamplified_ir (opo green pump un-shuttered)

@80uW pump trans, amplified = 0.0198 (at start, 0.0196 at end) --> NLG 0.0198/0.0014 ~ 14

@100uW pump trans, amplified = 0.0046 (at start, 0.0458 at end) --> NLG 0.046/0.0014 = 33

@120uW pump trans, amplified = 0.16 --> NLG 0.16/0.0014 = 114

@60uW pump trans, amplified = 0.011 (at start, 0.0107 at end) --> NLG = 7.86

@40uW pump trans, amplified = 0.0059 (at start, 0.0059 at end) --> NLG = 4.2

@20uW pump trans, amplified = 0.0034 (at start, --- at end) --> NLG = 2.4

trace reference opo_green_trans
(uW)
NLG SQZ dB CLF RF6 demod angles (+)
LO shot noise @ 1.106 mA, -136.3 dB 10 80 14    
Mean SQZ 11     +13  
SQZ 12     -8.0 162.0
ASQZ 13     +16 245.44
NLG=33   100 33    
Mean SQZ 14     +16.7  
SQZ 15     -8.0 170.5
ASQZ 16     +19.9 237.85
NLG = 114   120 114    
Mean SQZ 17     +22.5  
SQZ 19     -8.0 177.98
ASQZ 18     +25.6 230.13
NLG = 7.9   60 7.9    
Mean SQZ 20     +9.7  
SQZ 21     -7.7 154.28
ASQZ 22     +12.4 253.83
NLG = 4.2   40 4.2    
LO SN check 4       ~0.1dB lower?
Mean SQZ 23     +6.8  
SQZ 24     -6.3 140.6
ASQZ 25     +9.6 262.64
NLG = 2.4   20 2.4    
Mean SQZ 26     +3.8  
SQZ 27     -4.8 135.45
ASQZ 28     +6.3 -100.5
LO shot noise @ 1.06 mA, 29        

All measurements had PUMP_ISS engaged throughout; we manually tuned the ISS setpoint for different NLGs. For low NLG (20uW trans) we manually engaged ISS. LO power (shot noise) drifted ~5% over the measurement, see trends.

NLG Sweep Procedure:

DTT saved in $(userapps)/sqz/h1/Templates/dtt/HD_SQZ/HD_SQZ_8dB_101823_NLGsweep.xml

Images attached to this report
Non-image files attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 11:06, Thursday 19 October 2023 (73581)

Using Dhruva's nice plotting code for NLG sweeps from LHO:67242, here are some plots of squeezing vs. NLG, and calibrating the OPO lasing threshold and various green powers at this new crystal spot. Data & updated plotting code attached.

NLG sweep data summary here:

SHG Launched (mW) OPO Green Refl (mW) OPO Green Trans (uW)  NLG 
 
Mean SQZ (dB) SQZ (dB) Anti-SQZ (dB)
10.8 1 80 14 13 -8.0 16
13.3 1.3 100 33 16.7 -8.0 19.9
16 1.5 120 114 22.5 -8.0 25.6
8.4 0.8 60 7.9 9.7 -7.7 12.4
5.9 0.6 40 4.2 6.8 -6.3 9.6
3.5 0.4 20 2.4 3.8 -4.8 6.3

To-do: Look into the fits of loss & technical noise.

  • Technical noise here fits to -8.8dB below LO shot noise @ 1.1mA (-136dB, dark noise @ -159dB) -- check this is correct/real. For reference with the old homodyne PD in Feb 2023, LHO:67223 with 5.5dB sqz, technical noise fit to -9.6dB below LO shot noise @ 1.08mA (calibration suspect, HD_DIFF_DC_OUT was around -134-135dB, dark noise @ - 157dB). Would be interesting if this about -10dB technical noise below shot noise on the homodyne is real technical noise injection from the squeezer, and/or if it can be improved with upcoming squeezer laser PMC installation.
  • Loss estimate from NLG sweep - our homodyne loss budget accounts for 13% losses (after scaling back all over-estimations of losses), but this sweep fits to 7% losses, so we should resolve this over-estimated loss discrepancy.
Non-image files attached to this comment
michael.zucker@LIGO.ORG - 14:58, Thursday 19 October 2023 (73599)

Outstanding work, well done!

victoriaa.xu@LIGO.ORG - 15:29, Monday 20 November 2023 (74319)

Attached here is a re-fitting of this homodyne NLG sweep, which fits [loss, phase noise, technical noise] to measured SQZ+ASQZ, given measured NLG. It also shows the calculated loss from measured mean-sqz and NLG (which relies on accurate calibration of NLG --> generated SQZ dB). The same fitting was done for NLG sweeps on DARM the following week LHO:73747.

The previous anlaysis was fitting [phase noise, technical noise] using the loss calcuated from mean-squeezing. Compared to the earlier analysis, I think these fits here are closer. 

We budgeted 13% HD loss for this homodyne measurement:  1 - [0.985(opo) * 0.96(ham7) * 0.98(sqzt7 optics) * 0.977(HD PD QE) * 0.96((visibility~98%)**2)] = 13%.

This fit to the NLG HD sweep suggests ~11% homodyne loss, 7 mrad rms phase noise, with technical noise about -10 dB below 1.1 mW LO shot noise. Note HD dark noise is -22 dB below shot noise, suggesting mystery technical noise on the homodyne.

Images attached to this comment