Displaying report 1-1 of 1.
Reports until 20:14, Tuesday 24 October 2023
H1 General
anthony.sanchez@LIGO.ORG - posted 20:14, Tuesday 24 October 2023 (73724)
Tuesday EVE Mid Shift Report

Took ISC_LOCK to INITIAL_ALINMENT and found almost no light on both arms.
I then Exited INITIAL_ALIGNMENT, and went to Manual Green Arms.
Set Increase flashes on Y and Manualed X all the way up to 80%
Increase flashes couldn't get above 55ish% on Yarm so I then touched up Y arm and got it to 1.
Sent ISC_LOCK back to I.A.
Once Initial Alignment completed the lock went well up until ISC_LOCK got to  
PREP_DC_READOUT_TRANSITION where we got stuck due to a change in the OMC settings which were resolved by Vlad and Jenne. See comment 73720 https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=73718

We continued up to POWER_10W and caught a lockloss.
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1382229688
picture taken before 17:45


Relocking had another lockloss at Find IR.

01:10 UTC EY temp is still high but the slope of the temp trend seems to be trying to turn around.
H0:FMC-EY_VEA_AVTEMP_DEGF = 65.50
2:26 UTc Vicky Gets back to us and says that the SQZr Looks good.

Used the following command to mananged the blends under SEI_CONF
cdsutils write H1:GRD-ISI_ITMY_ST1_BLND_MANAGER SEI_CONF
cdsutils write H1:GRD-ISI_ETMX_ST1_BLND_MANAGER SEI_CONF

Finally got into NOMINAL_LOW_NOISE at 2:34 UTC and started sorting through SDF Diffs, then went to NLN_CAL_MEAS to run the following command at 2:42 UTC :
hwinj detchar safety --run
and back to NOMINAL_LOW_NOISE And OBSERVING at 2:51 UTC

 

 

Displaying report 1-1 of 1.