Displaying report 1-1 of 1.
Reports until 12:34, Wednesday 25 October 2023
H1 CAL
louis.dartez@LIGO.ORG - posted 12:34, Wednesday 25 October 2023 (73736)
Corrections to modeled response function since start of O4a
Here is a list of the _known_ corrections to the systematic error that need to be accounted for when regenerating the O4a uncertainty budgets.


IssueStartEndStatus
3.2 kHz ESD Driver Pole in CAL-CS Model (LHO:72043)2023/04/26 (ER15)2023/08/08 00:15 UTCRough fit exists. LHO:71787
Spoiled TDCFs due to noise in Kappa_UIM (LHO:71790, LHO:71846) 2023/07/20 2023/08/03 In progress. Pending correction file.
Thermalization effects seen in sensing function at 75W ER15 2023/06/21 In progress by Cal group.
Thermalization effects seen in sensing function at 60W 2023/06/21 Now In progress by Cal group.
There may be other corrections that are needed by additional unmodeled effects including OM2 hot/cold, drifting Kappa_T, and Pcal error that hasn't been accounted for (if any). Useful reference: Record of Real-time Calibration Pipeline Parameter Changes: O4 (and ER15) -- H1 (DCC T2300297)