Displaying report 1-1 of 1.
Reports until 15:47, Friday 27 October 2023
H1 ISC
camilla.compton@LIGO.ORG - posted 15:47, Friday 27 October 2023 - last comment - 15:04, Tuesday 14 November 2023(73773)
MICH FF Remeasured but not Changed

Took MICH FF measurements following instructions in lsc/h1/scripts/feedforward/README.md but running MICHFF_excitation_ETMYpum.xml. Data saved in lsc/h1/scripts/feedforward/ as .txt files. Last tuned MICH in 73420

I saved a new filter in FM6 as 27-10-23-b (red trace) but it made the MICH coupling worse between 20 and 80Hz so we left the original (pink trace). We could try to re-fit the data to load in Wednesday's commissioning period.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 14:49, Sunday 29 October 2023 (73821)

I re-fit this data and think i have a better filter saved (not loaded) in FM3 as "27-10-23-a". We could try this during a commissioning period this week if we wanted to try to further improve MICH.

Plot attached of proposed filter FM3 (red) with current filter FM5 (blue REF0) and Friday's failed filter FM6 (green REF1). 
We would like to be close to FM5 (blue) 20-80Hz close to FM6 (green) above and below this. Proposed filter does have increased gain at low frequencies.
Images attached to this comment
camilla.compton@LIGO.ORG - 15:04, Tuesday 14 November 2023 (74206)

Tried this FM3 FF 2023/11/14 16:04:30UTC to 16:06:30UTC. It did not cause a lockloss. I did not run the MICH comparism plot but DARM looked slightly worse. Plot attached.

From 16:07:05, I tried FM6 which is the currently installed MICH FF (FM5) without the 17.7Hz feature 74139.

Images attached to this comment
naoki.aritomi@LIGO.ORG - 13:57, Friday 03 November 2023 (73962)

I tried to test a new MICH FF FM3 Camilla made. First I measured the current MICH FF FM5 as shown in the attached figure. The pink and black curves are the current MICH FF FM5 on 20231027 and 20231103, respectively. The current MICH FF gets worse between 30-80 Hz in a week. The MICH FF on 20231103 was measured after 6.5 hours into lock. Then I ramped the MICH FF gain to 0 and turned off FM5 and turned on FM3. After I ramped the MICH FF gain to 1, a lockloss happened immediately.

Images attached to this comment
camilla.compton@LIGO.ORG - 11:13, Monday 06 November 2023 (74021)

Sorry that this caused the 1383077917 lockloss.

Unsure why this FM3 would be unstable. Lockloss occurred 10 seconds after MICHFF had finished ramping on (13s - 3sec ramp time). FM3 MICH_FF looks to be outputting ~ factor of 2 higher than the current FM5 filter. Don't see any obvious instabilities in the 10seconds before the lockloss.

LSC and ASC plots attached. I wonder if the lockloss was just badly timed. We could attempt to repeat this before our Tuesday Maintenance period.

Images attached to this comment
Displaying report 1-1 of 1.