H1 AOS (DetChar)
robert.schofield@LIGO.ORG - posted 18:46, Sunday 05 November 2023 (74013)
Peak in DARM from CHWP-2 at EX reduced or eliminated by changing pump frequency from 52 Hz to 59 Hz

Genevieve, Lance, Tyler, Robert

In August, we determined that the 52 Hz peak in DARM could be dramatically reduced or eliminated by turning off the chilled water pump CHWP-2 at EX (72331), which was running at 52 Hz. On Friday I decided to check if the water pump was lining up in frequency with a mechanical resonance that made the coupling worse. So I used the variable frequency drive to move the pump frequency around. The figure shows that I found three frequencies where the water pump did not produce an obvious peak in DARM and one where it produced a bigger peak in DARM.

After consulting with Tyler, I have left the frequency at 58 Hz (98% on the FMCS screen). Of course whatever resonance was associated with the peak in DARM at 52 Hz will still be excided by non-pump seismic motion, so the 52 Hz peak may not go away completely. I may move the pump frequency around some more this week to better understand the coupling.