Reports until 16:28, Friday 09 August 2013
LHO VE
kyle.ryan@LIGO.ORG - posted 16:28, Friday 09 August 2013 (7405)
Manual CP1 level control passed to CDS PID @ 1230 hrs. local
Kyle 

Had valved-out inst. air to LLCV this morning while manually filling CP1 -> Hand over to PID control re-exposed known quirk with CP1's pneumatic actuator control box, that being a gas-column oscillation -> I adjusted (cw) the damping jet until the oscillations stopped -> Control behavior may be different as a result 

I will monitor from home