SQZ Laser Frequency/Phase Stabilization (SQZ-LO) Loop Servo Fast Path's VCO Calibration Turned ON
S. Dwyer, J. Kissel, D. Sigg

While finding out the best effective frequency noise estimate on the SQZ side of things, we found a VCO calibration in FM3 available on the fast path of the SQZ LO frequency servo (H1:SQZ-LO_SERVO_CTRL), which turns that out-of-loop, pick-off path into a diagnostic effective frequency noise estimate (in lambda = 1064e-09 [m] "red" light). See attached screenshot for how to find this filter bank.

While we can't quickly find any documentation/aLOG on this calibration, it makes sense to us that it has the same frequency response (a zero at ~40 Hz, and a pole at ~1.6 Hz) and roughly the same gain as the primary CARM / IMC Common Mode board servo's fast path (~237000 [Hz/V]).


    IMC VCO calibration     FM3 "VCO"   zpk([40],[1.6],237000,"n")         LHO:65175 (Updated Sept 2022)
    SQZ LO VCO calibration  FM3 "VCO"   zpk([39.72],[1.5292],260000,"n")   (Installed O3 era, we think)

I've accepted the VCO, FM3 filter as ON in the H1SQZ safe.snap SDF, and confirmed that it's either ignored or by default *also* saved to the OBSERVE.snap. 
Images attached to this report