H1 AOS
sheila.dwyer@LIGO.ORG - posted 15:34, Friday 17 November 2023 (74264)
DARM OLG measurement with ETMX A2L changes

There is another alog from Louis coming soon, but we have been looking at the PUM crossover and see that it isn't well described by the pyDARM model.  Since the crossover measurement has low coherence and can cause locklosses, today we did some investigation using the DARM OLG measurement.  The attached plot shows a DARM OLG measured with our normal A2L decoupling (P2L = 4, Y2L = 4.4) and with half those values (P2L = 2, Y2L = 2.2), you can see that the DARM OLG is different between 20-5 Hz.  

 

Images attached to this report