Displaying report 1-1 of 1.
Reports until 12:59, Friday 24 November 2023
H1 PSL
austin.jennings@LIGO.ORG - posted 12:59, Friday 24 November 2023 (74397)
PSL Status Report

Closes 26219, last done on November 19th


Laser Status:
    NPRO output power is 1.813W (nominal ~2W)
    AMP1 output power is 68.05W (nominal ~70W)
    AMP2 output power is 136.8W (nominal 135-140W)
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN

PMC:
    It has been locked 3 days, 3 hr 39 minutes
    Reflected power = 16.07W
    Transmitted power = 109.5W
    PowerSum = 125.6W

FSS:
    It has been locked for 0 days 8 hr and 7 min
    TPD[V] = 0.9551V

ISS:
    The diffracted power is around 2.9%
    Last saturation event was 0 days 8 hours and 8 minutes ago


Possible Issues: None