terra.hardwick@LIGO.ORG - posted 18:22, Thursday 15 August 2013 - last comment - 08:13, Friday 16 August 2013(7456)
Magnetometer tripod mount prototype tested
We've tested a new tripod mounting system for the magnetometers with aims to decrease artifacts from motion of the magnetometer on the tripod. The iLIGO mounting scheme resulted in peaks in the magnetometer spectra that were not from real magnetic fields but from the magnetometer moving in the earth's field at tripod resonances. We tap tested both tripods to study their resonances and damping. The flagpole resonance of the current tripod is at 7.9 Hz. The new system, designed at Cal State Fullerton, is more damped and rigid. 

Attached is the power spectrum of both setups, side to side (see photo). The red trace is the magnetometer mounted on the current tripod, the blue trace is the magnetometer mounted on the Fullerton prototype tripod. The flagpole resonance of current tripod is seen at 7.9 Hz, the one place the blue Fullerton trace does not follow. Most other peaks seen are from nearby cleanroom fans and the expected 60 Hz. 

Terra, Robert
Images attached to this report
Non-image files attached to this report
Comments related to this report
john.worden@LIGO.ORG - 21:30, Thursday 15 August 2013 (7462)

Two tripods in the photo - is the orange one the current or the Fullerton?

terra.hardwick@LIGO.ORG - 08:13, Friday 16 August 2013 (7464)
Orange is the Fullerton. 