Since the PSL rotation stage has not been calibrated since July, I took the time to do it again this morning during maintenance. The process was as follows:
- Took IMC offline and paused LASER_PWR guardian
- Ran the script /opt/rtcds/userapps/release/psl/h1/scripts/RotationStage/MoveRotationStage.py to create PSL_rotStageCalib_5Dec2023.txt (attached)
- Changed PSLvals in CalibRotStage.m for correct file name then ran to create calibration curve (attached) and get new fit values (in table)
- Input new fit values into calibration and accepted new values in SDF (screenshot attached)
- Searched for HOME on rotation stage
- Resumed LASER_PWR guardian to go back to 2W input power
|
Power in (W) |
D |
B (Minimum power angle) |
C (Minimum power) |
Old Values |
100.917 |
1.990 |
-24.827 |
0.000 |
New Values |
102.9407 |
1.990 |
-24.803 |
0.000
|