We installed the new MICHFF FM8 filter that Gabriele fit 74602, 74595. This looks to do a great job suppressing the MICH excitation, see pink trace attached. A little less suppression 150-200Hz so we'll want to run some coherences with NLN data to see if we care about MICH in this region. Saved in safe and observe sdf and ISC_LOCK.
Left the MICHFF off for 5 minutes 22:05:45 to 22:10:45UTC.
Tagging CAL, this may have an effect on the 17.6Hz line 74259