I ran a Bruco for yesterday after commissioning time ended: 2023/12/07 01:07:00 UTC bruco_1385946438_STRAIN and the hour before commissioning period started: 2023/12/06 18:17:30 UTC bruco_1385921868_STRAIN The IFO had been locked since 2023/12/06 09:00UTC, so for 9 and 16 hours before measurements.
MICH coherence looks much better after compared to before the new FF was installed 74642. Can also compare to October's 73701 Bruco. There is still coherence around the 17Hz BS mode 74259, maybe this is expected.
Other ones to note: CHARD_Y, CHARD_P, PRCL, SRCL
Ran in LDAS using commad: python -m bruco --ifo=H1 --channel=GDS-CALIB_STRAIN_NOLINES --gpsb=1385921868 --length=3600 --outfs=1024 --fres=0.1 --dir=/home/camilla.compton/public_html/bruco_1385921868_STRAIN --top=100 --webtop=20 --plot=html --nproc=20 --xlim=1:1000 --excluded=/home/elenna.capote/bruco-excluded/lho_DARM_excluded.txt