Overview of combs in recent monthly Fscan spectra

This is a big-picture overview of combs in recent months.

Data and methods

"Relative priority" is a new metric documented in G2302324 (Athena Baches, Abby Wang, and Nick Baker). It is not finalized-- in fact these values differ slightly from the version in the linked presentation due to bug fixes-- and should be understood as (a) having arbitrary units, (b) reflecting the general importance of spectra artifacts to CW searches, and (b) being useful only for comparing artifacts in similar data sets-- for example, it should not be used to compare artifacts from monthly spectra and ones from weekly spectra. In this case it is used to compare and rank combs across monthly spectra.

Monthly August, October, and November Fscan spectra and their auto-generated lines lists were used. I skipped September because I still need to troubleshoot some issues with the lines list for that month.

Main messages

Details

Key

Color-coding and letter labels convey identical information.

A
9.47-8 Hz comb triplet. (70219) Duplicate entries likely due to frequency shifts, which have been observed for this comb (72323) Cause unknown.
B
1.66 Hz comb related to OM2 heater driver and variations which appeared during testing in October. Now mitigated. (72967)
C
near-1Hz and near-5Hz combs recently associated with HWS (74614, 74617)
D
near-30 Hz comb. Cause unknown.
E
set of near-100Hz combs with various offsets. Cause unknown
no tag
Entries that are not persistent over multiple months.

August

Ranking Comb spacing Comb offset Relative priority Label
1 9.47425 0 46  A
2 9.47539 0 45  A
3 1.6611234 0 44  B
4 9.480545 9.48 39  A
5 4.9842174 0 22  C
6 29.969515 0 17  D
7 99.998456 70.029 5  E
8 6.64457 1.659 5  
9 99.99846 29.9695 5  E
10 99.99864 0 5  E
11 99.99844 89.909 2  E
12 99.99846 0 2  E
13 199.9969 10.09 1  E

October

Ranking Comb spacing Comb offset Relative priority Label
1 9.475416 0 43  A
2 9.474265 0 43  A
3 9.48057 0 42  A
4 9.480574 0 39  A
5 0.996787 0 32  C
6 4.98425 0 30  C
7 1.10864403 0 23  B
8 29.969514 0 15  D
9 1 0 6  
10 99.99846 29.9694 5  E
11 99.998644 0 5  E
12 99.99846 70.029 4  E
13 99.99845 10.09 2  E
14 99.99845 0 2  E

November

Ranking Comb spacing Comb offset Relative priority Label
1 9.47429 0 47  A
2 9.4754423 0 34  A
3 9.480612 0.0003 34  A
4 9.475456 9.4748 33  A
5 0.996785 0 32  C
6 4.984252 0 29  C
7 29.96951 0 15  D
8 28.44192 9.48 8  
9 0.12332754 0.1166 8  
10 0.1233334 0.0614 7  
11 47.4031 28.4417 6  
12 1 0 6  
13 99.998458 70.029 4  E
14 99.99846 29.9694 4  E
15 99.99844 0 2  E
16 99.99844 89.9088 2  E
17 8.426278 5.5594 1  
Displaying report 1-1 of 1.