Sat CP1 Fill, manually stopped overfill

Sat Dec 16 10:07:32 2023 ALERT: Fill done (errors) in 7min 28sec

Gerardo manually stopped the overfill after it became clear the TCs were not going to get down to the -80C trip temps (bottomed out at -77C and -71C).

TCs started with positive readings (+10C, +14C) outside temp was 0.4C at the time in very foggy conditions.

Images attached to this report