Camilla, Ansel
At 17:46UTC, Camilla changed ITMX and ITMY HWS to 'sem 3', which is external trigger mode. This seemed to work, and stopped the code taking photos.
It appears that the HWS-associated combs are gone in the magnetometer witness channel after this change. Pre/post 1-hour spectra attached. (All known HWS-associated combs overlaid, just to check-- the near-7Hz is the one actually present in the "pre" spectrum which corresponds to current HWS sync frequency settings.)
Nice work Camilla and Ansel! Let's hope this solves the problem for good.
At 17:38UTC I tunred on the camera/CLink and restarted the HWS EX code which had been off since 74738.
At 17:46UTC I changed ITMX and ITMY to 'sem 3' and left them in this configuration so that they are not taking HWS data now. Now we need to write a script to read H1:GRD-ISC_LOCK_STATE_N state and adjust the camera mode to be sem 2 when we are in states < 580 (locking) and sem 3 when > 580 (locked). This is not trivial as camera computers are separate from EPICS.