Wed Dec 27 10:07:13 2023 INFO: Fill completed in 7min 9secs
Gerardo, Richard, Dave:
Gerardo confirmed a good fill curbside.
Gerardo is removing the ice ball at the end of the discharge line which was causing erroneous readings. Currently TC-B is not giving a good value (~1300C). Today's fill ran with TC-A only.
I've temporarily bypassed TC-B's cell phone alarms while Gerardo is working on it.
Bypass will expire:
Thu Dec 28 10:19:24 AM PST 2023
For channel(s):
H0:VAC-LY_CP1_TE102B_DISCHARGE_TEMP_DEGC
H0:VAC-LY_CP1_TE102B_DISCHARGE_TEMP_DEGC_ERROR
Thermocouple wire for TC-B were spliced together with wire nuts, then some e-tape was used to cover both wire nuts. TC-B is now giving good temperature readings
Filed an FRS ticket to replace thermocouple, 30032