After the glitches last night, I went to HAM3 to do the usual fixes. I powered off the corner 2 CPS, unplugged the boards in the satellite chassis at the chamber, put it all back together and powered everything back on. The CPS haven't glitched since then, so seems like maybe things are fine now.
The 65-100 hz blrms are a good witness glitches for a couple hours before they started tripping the HAM3 ISI. Attached trend shows about 5 hours before the first HAM3 trip on Monday. The top row are the raw H2 and V2 CPS in counts, the middle row is the watchdog state, and the bottom row are the 65-100hz blrms for the corner 2 and corner 3 blrms. The H2 and V2 cps start seeing glitches that don't trip the ISI about 3 hours before the first trip, these glitches don't really show up in the in the corner 3 CPS either. These glitches also don't coincide with locklosses, if the ISI doesn't trip. Under normal circumstances, these blrms are well below 10 nm, the first few glitches are up to 600 nm, but a glitch of ~1000nm causes the ISI to trip. There haven't been any glitches since I touched the CPS yesterday, so I think we are in the clear for now.
I'm still not sure of the right way to alarm on this, but some sort of days-ish timeseries trend when ISI trips on CPS would probably be a good place to start.