Displaying report 1-1 of 1.
Reports until 13:33, Saturday 06 January 2024
H1 PSL (PSL)
ibrahim.abouelfettouh@LIGO.ORG - posted 13:33, Saturday 06 January 2024 (75219)
PSL Weekly Report - Weekly FAMIS 26225

Closes FAMIS 26225


Laser Status:
    NPRO output power is 1.819W (nominal ~2W)
    AMP1 output power is 67.9W (nominal ~70W)
    AMP2 output power is 137.5W (nominal 135-140W)
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN

PMC:
    It has been locked 30 days, 23 hr 15 minutes
    Reflected power = 16.9W
    Transmitted power = 109.4W
    PowerSum = 126.3W

FSS:
    It has been locked for 0 days 0 hr and 30 min
    TPD[V] = 0.9213V

ISS:
    The diffracted power is around 2.2%
    Last saturation event was 0 days 1 hours and 43 minutes ago


Possible Issues: None

Displaying report 1-1 of 1.