Camera issues on h1digivideo1

Erik, Dave, Patrick, TJ:

At 22:55 Tue PST the PR3 camera stopped updating. This morning I power cycled the camera and restarted the server process on h1digivideo1.

Soon after the AS AIR camera started blue-screen flashing on clients viewers. Each blue flash was short (less than a second),  the EPICS values did not show any flat-lining. The flash rate was random and about every 5 minutes.

We found nothing untoward on h1digivideo1.

For a first try at fixing this we restarted the server process on h1digivideo1 for this camer (h1cam16). This did not fix it.

For a second try we power cycled the camera via POE control and restarted the server process, again this did not fix it.

Then more cameras on h1digivideo1 started blue screen flashing, so at 13:37 we did a soft reboot of h1digivideo1.

Comments related to this report
david.barker@LIGO.ORG - 14:18, Wednesday 10 January 2024 (75307)

Did not fix it, flash rate is several per minute. Opened FRS30147