Julian, Camilla. WP 11614
With 4mW seed beam injected, we touched the quater and half waveplates in the SEED path on SQZT0 to minimize H1:SQZ-CLF_FIBR_REJECTED_DC_POWER. Plot attached. Reduced HAM7 rejected from 0.32mW to 0.12mW (we've not previously got this to 0 65063).
Throughput to SQZT7 is 1.2mW/4mW = 30%. We expect >40% and we have previously got only get 50%-75% though the fiber coupler switch only (not inc HAM7) 65392 65445.