Displaying report 1-1 of 1.
Reports until 10:22, Monday 26 August 2013
H1 PSL
christina.bogan@LIGO.ORG - posted 10:22, Monday 26 August 2013 (7535)
Restarted Laser after power outage
[Peter, Christina]

We restarted the laser and found out that after the power outage the current limit of the HV units was set to 0 (I recall that is a known problem). We set it to 200mA and the laser locked immediately. The PSL came back without any other issues.