Closes FAMIS 26227
Laser Status:
NPRO output power is 1.835W (nominal ~2W)
AMP1 output power is 68.84W (nominal ~70W)
AMP2 output power is 140.0W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PMC:
It has been locked 1 days, 20 hr 55 minutes
Reflected power = 18.8W
Transmitted power = 108.8W
PowerSum = 127.6W
FSS:
It has been locked for 3 days 0 hr and 28 min
TPD[V] = 0.3952V
ISS:
The diffracted power is around 1.7%
Last saturation event was 3 days 1 hours and 0 minutes ago
Possible Issues:
PMC reflected power is high
FSS TPD is low
ISS diffracted power is low