H1 PEM (PEM)
marc.pirello@LIGO.ORG - posted 16:29, Wednesday 07 February 2024 (75772)
MY Accelerometer Chassis Installed

Continued from previous ALOG75713

Replaced Meggit (Endevco) Accelerometer Power Conditioner with LIGO Accelerometer Power Conditioner. WP11653
S2300076 installed, all signals attached and powered up at 4:03pm PST.

We did have inrush current blow a fuse on the line, we replaced the 10A fast blow with a 3A slow blow fuse and this solved the issue.

The IO chassis did trip when we switched the accelerometer chassis on, I worked with EJ and Jonathan to restore the connection.

To Do:
MX power adapter cable install and signals transferred from Endevco boxes to the LIGO units.