Reports until 12:20, Monday 12 February 2024
H1 CDS (AWC, ISC, SQZ, SUS)
filiberto.clara@LIGO.ORG - posted 12:20, Monday 12 February 2024 (75827)
HAM7 Ground Loop Checks

D2000202 A+ SUS HAM 7 System Wiring Diagram
T1200131 Grounding and Shielding at LIGO
D1900511 ISC/SQZ Wiring Diagram

Ground loop checks for HAM7 complete. No issues to report. Some cables previously modified per E2100504. Other don’t have Pin 13 and shield tied per T1200131. See alog 65129 for more details.

 A list of cable tested:

Description Cable Location  
FC1 TOP SUS_FC1_001 SUS-R4  
FC1 TOP SUS_FC1_002 SUS-R4  
FC1 MIDDLE SUS_FC1_003 SUS-R4  
FC1 Bottom SUS_FC1_004 SUS-R4  
VOPO SUS_OPO_001 SUS-R4  
VOPO SUS_OPO_002 SUS-R4  
ZM1-M1 SUS_ZM1_001 SUS-R4  
ZM1-M2 Dither SUS_ZM1_006 SUS-M1  
ZM2-M1 SUS_ZM2_001 SUS-R4  
ZM2-M2 PSAMS AWC_ZM2_006 SQZ-R1  
ZM3-M1 SUS_ZM3_001 SUS-R4  
ZM3 M2 Dither SUS_ZM3_006 SUS-M1  
ZM4-M1 SUS_ZM4_001 SUS-R4  
ZM4-M2 PSAMS AWC_ZM4_006 SQZ-R1  
ZM5-M1 SUS_SQ_002 SUS-R4  
ZM5-M2 PSAMS AWC_ZM5_006 SQZ-R1  
Oven Trans/Cav Piezo ISC_SQ_303 SQZ-R2  
SQZ Thermistor & TEC ISC_SQ_304 SQZ-R2  
SQZ DCPD & LENS Trans ISC_SQ_305 SQZ-R2  
Picomotor ISC_SQ_470 ISCT7  
Picomotor ISC-SQ_471 ISCT7  
SQZ-FC_QPD_A ISC_SQ_479 SQZ-R1  
SQZ-FC_QPD_B ISC_SQ_480 SQZ-R1  
OFI2 ISC_SQ_482 SQZ-R2  
SF1 ISC_SQ_483 SQZ-R2  
SF2 ISC_SQ_484 SQZ-R2