EtherCAT Squeezer Chassis 5

WP 11707

SQZ Chassis 5 modified to add missing ground connection for binary input and output signals. Ground cable connected from EL2124 (Middle Rail, Terminal 2, Signal labeled "GND") to Adapter 6 (Pin 19).

Serial Number S2300256

F. Clara, D. Sigg