Reports until 13:07, Wednesday 21 February 2024
H1 AOS
filiberto.clara@LIGO.ORG - posted 13:07, Wednesday 21 February 2024 (75915)
EtherCAT Squeezer Chassis 5

WP 11707

SQZ Chassis 5 modified to add missing ground connection for binary input and output signals. Ground cable connected from EL2124 (Middle Rail, Terminal 2, Signal labeled "GND") to Adapter 6 (Pin 19).

Serial Number S2300256

F. Clara, D. Sigg