Both access ports on -Y door for HAM8 were leak tested. The helium background registered at the leak detector was of 4.6x10^-10 Torr-l/s before starting the test. Both joints were sprayed with helium and no signal above the background was registered by the leak detector. Final background registered by the leak detector once the leak test was done was of 4.0x10^-10 Torr-l/s.